Wechat QR code

美伦医疗微信公众号二维码.jpg

TEL:400-654-1200

Million Medical 

Stock code:871562

Phone:

400-654-1200

What do you want to know?

监护仪厂家,脑电图机,美伦,美伦医疗电子有限公司
Current position: Homepage >> News >> Company news

Maintenance and maintenance of electromyograph

The number of clicks:
Release date: 2018-11-19 00:00:00
Source: Henan Meilun Medical Electronics Co., Ltd.
share it

Maintenance and maintenance of electromyograph


(1) Storage and routine maintenance

Instruments should be placed in a ventilated and dry place. Air conditioning must be provided in the room. Do not place the instrument under the window or beside the heating to avoid showering, heating and dampness. When the instrument is out of use for more than ten days, it should be electrified every other week for one hour each time to prevent corrosion and damage.Meilun

It is better to use special power supply system, such as voltage stabilization device. Only when the voltage stabilization device reaches the required voltage of the instrument, can the power switch be turned on, so as not to damage the instrument by too high or too low voltage.

After use, the knobs and switches are restored to their original position, and the electrodes connected to the preamplifier and stimulator are pulled out, then the power switch is turned off.

(2) Cleaning and disinfection

Before cleaning or disinfecting the surface of the instrument, disconnect the power cord of the instrument.

The external surface (physiological amplifier, main control box, current stimulator, audio and video stimulator box, display, printer, computer host, etc.) should be cleaned regularly, and the wet cloth without hair removal should be used for cleaning.

Always dilute the solution as recommended by the manufacturer.

Always wipe all cleaning solutions with a dry cloth after cleaning.

Meilun

Never use waxy cleaning substances.

Never sprinkle or spray water or any clean solution onto the equipment, or allow the liquid to flow behind the switch, into the connector or into any vent in the equipment.

Never use the following detergents: any kind of detergent or solvent, acetone, methyl ketone, alcohol-containing detergent.

To clean the display screen, use clean soft cloth and wet it with detergent. Never spray detergents directly onto the screen, nor use alcohol or medical disinfectants.

Cautions for wire cleaning of connecting electrodes: Do not use acetone or ketone solution for cleaning; Do not use high-pressure sterilizer or steam cleaner; Do not soak patient wire; Patient wire can be cleaned with warm, moist cloth and mild soapy water, or wiped with isopropanol.Meilun

Notes for headphone cleaning: Do not use acetone or ketone solution for cleaning; Do not use high-pressure sterilizer or steam cleaner; Do not soak headphones; Before and after each use of headphones, use 75% medical alcohol to clean the part of the headphone that contacts the human ear; Other parts of the headphones can be cleaned with warm, damp cloth and mild soapy water, or Wipe with isopropanol. Meilun

Notes for cleaning the signal collection electrode: After using the disc silver electrode, remove the conductive paste from the surface of the electrode plate, wipe the electrode plate with a absorbent cotton cloth soaked in alcohol, and then dry it; if the disposable electrode is used, it should be destroyed according to the relevant national standards.

Precautions for cleaning electrodes of electroencephalograph by electric current stimulation: before and after each use, 75% medical alcohol must be used to clean the silver electrodes contacted with human body; other parts can be cleaned with warm, moist cloth and mild soapy water, or wiped with isopropanol; do not use any abrasives to clean electrodes, any scratches on the surface of electrodes can cause graphics. The record is inaccurate.


Maintenance and maintenance of electromyograph