How do I clean and maintain the multi-parameter monitor?

First, clean:

1, the use of soft cotton and other cleaning equipment;

2, the host and the power box can use soap, amino or ethanol-based cleaning materials;

3, reusable ECG lead line, blood pressure sensor, blood oxygen probe, body temperature probe can use soap or amino cleaning materials.

4, every 1-2 months to clean the host vents filter, every six months to 1 year to open the host shell, the machine for dust, clean the host surface dirt.

5, the ambient temperature of the equipment is 5-40 ℃, relative humidity less than 80%, the working environment requires dry and well ventilated, in order to facilitate heat and moisture. Such as long-term do not use, should be regularly boot, so that heating to achieve moisture-proof purpose.

6, to avoid contact with acid and other equipment, corrosive gases and liquids, so as not to damage. Do not pour or spray any cleaning solution to the equipment to prevent liquid from flowing into the cabinet, switch, interface and vent.

Second, sterilization:

1, the host and the power box should use ethanol, acetaldehyde cleaning agent;

监护仪厂家

2, ECG lead line should use acetaldehyde cleaning agent;

3, reusable cuff, oxygen probe, body temperature probe should use ethanol, acetaldehyde cleaning agent.

Third, disinfection:

Monitor manufacturers on a regular basis on the basis of cleaning, if necessary, do disinfection. Disinfection must be strictly in accordance with the norms of the hospital, with particular attention to safety. Generally use EtO gas disinfection.

Fourth, maintenance:

1, regularly check the instrument, lead line, sensors and accessories;

2, according to the need for equipment cleaning;

3, once a year to conduct a safety test check;

4, once a year calibration NIBP and body temperature parameters;

5, an annual machine function test;

6, each boot repair, you must conduct a safety test check;

7, on the regular built-in rechargeable battery maintenance, every two months to optimize the battery.

Fifth, operational considerations

1, can not be folded, drag the cable and sensors, storage cables and sensors should be natural to complete the circular placement, can not fall, touch the sensor probe.

2, in order to prevent ECG waveform interference, ECG lead line requirements clean and smooth.

3, the monitor needs a long time continuous use, and do not need to monitor the oxygen saturation, the oxygen saturation monitoring function can be directly or directly remove the cable.


The above is from the cleaning, sterilization, disinfection, maintenance, operational considerations five aspects of the multi-parameter monitor were done in detail, usually do the equipment maintenance and maintenance work, to a certain extent, can reduce the equipment at the fault Probability, extend the service life of the equipment, reduce the cost of medical use, it is recommended that the hospital set up a special medical staff responsible for the cleaning of equipment.


How do I clean and maintain the multi-parameter monitor?