Wechat QR code

监护仪厂家,脑电图机,美伦,美伦医疗电子有限公司

TEL:400-654-1200

Meilun Medical

Stock code 871562

Phone:

400-654-1200

What do you want to know?

监护仪厂家,脑电图机,美伦,美伦医疗电子有限公司
Current position: Homepage >> News >> Company news

Maintenance of EEG machine

The number of clicks:
Release date: 2015-08-04 00:00:00
Source: Henan Meilun Medical Electronics Co., Ltd.
share it
The storage and use of the EEG machine should be relatively dry, clean, the temperature should not be too high or too low, the floor to prevent vibration, but also to prevent dust entering the instrument. The cloth should cover up when not in use. EEG instrument must be responsible for the custody and use of. In addition, we should note the following:
, EEG machine if long-term cessation of use must periodically energized, to ward off moisture and prevent rust, general monthly pass electric one to two times, each time through time an hour or so.
II, the use of ink is special for the use of ink, not the use of other writing ink, in case of record pen plug. If the ink tank leakage phenomenon, should immediately stop using, timely treatment. Special attention should not be made to ink into the recorder, because it will have a bad effect on the performance of the recorder.
To pay attention to the protection of the recording pen, to obtain high frequency response, the pen is always made of light, but at the same time also caused by the intensity of the pen is not enough, easy damage. When used, to prevent overload of the signal to the recorder.
4, to try to make the baseline in the middle position, as found in circuit fault and other reasons make recording pen partial to one side should be promptly turn off the record guide switch because recording pen to near the recorder of the coil current, for a long time, the coil will overheat and damage.
Gymnasium, pay attention not to let the saline wet electrode wire, branching, instrument panel and, if salt water intrusion inside the junction, will reduce the insulation caused by AC interference and other ills.
We should always check the situation, chloride electrode, and a chloride treatment after a period of use.


Maintenance of EEG machine